‌Step‌ ‌Action‌ ‌Verification‌
‌1‌ Apply ‌0%‌ pressure (vent to atmosphere). Output = ‌4.00 mA ±0.02 mA‌.
‌2‌ Adjust ZERO screw/HART parameter until output stabilizes at 4.00 mA. –
‌3‌ Apply ‌100%‌ pressure (e.g., 10 bar). Output = ‌20.00 mA ±0.03 mA‌.
‌4‌ Adjust SPAN screw/HART parameter to achieve 20.00 mA. –
‌5‌ Repeat Steps 1–4 until both endpoints meet tolerance. Deviation ≤ ‌±0.05% FS‌.

Shipping & Delivery

  • Courier delivery

Our courier will deliver to the specified address

2-3 Days

From $20

  • DHL Courier delivery

DHL courier will deliver to the specified address

2-3 Days

From $40

  • Warranty 1 year
  • Free 30-Day returns

Description

Calibration Procedure for EL3020 Transmitter‌ ‌

‌1. Pre-Calibration Setup‌

‌Tools Required:‌

  • Reference pressure source (accuracy ≤ ±0.1% FS)
  • Digital multimeter (resolution 0.01 mA)
  • HART communicator / Field calibration software

‌Safety Precautions:‌

  • Isolate process pressure and vent the transmitter.
  • Verify ambient temperature: ‌15–25°C‌ (outside range requires compensation).


‌2. Zero & Span Adjustment‌

‌(Apply using calibration rig)‌

‌Step‌ ‌Action‌ ‌Verification‌
‌1‌ Apply ‌0%‌ pressure (vent to atmosphere). Output = ‌4.00 mA ±0.02 mA‌.
‌2‌ Adjust ZERO screw/HART parameter until output stabilizes at 4.00 mA. –
‌3‌ Apply ‌100%‌ pressure (e.g., 10 bar). Output = ‌20.00 mA ±0.03 mA‌.
‌4‌ Adjust SPAN screw/HART parameter to achieve 20.00 mA. –
‌5‌ Repeat Steps 1–4 until both endpoints meet tolerance. Deviation ≤ ‌±0.05% FS‌.

‌Note‌: For smart transmitters, use Endress+Hauser FieldCare or 375 HART Communicator to automate adjustments.


‌3. Linearity Verification‌

‌Test Points:‌ 0%, 25%, 50%, 75%, 100% of range.

  • Record output current at each point.
  • ‌Acceptance Criteria‌:Error=∣ActualmA−ExpectedmA∣SpanmA×100%≤0.1%

‌4. Post-Calibration Actions‌

  1. ‌Seal Adjustment Screws‌ with anti-tamper seal (if applicable).
  2. ‌Update Calibration Certificate‌ with:
    • Date, technician ID
    • Reference tool serial numbers
    • As-found/as-left data.
  3. ‌Functional Test‌: Simulate overpressure (110% FS) to confirm fault alarm activation (HI-HI limit).


Critical Notes‌

  • ‌Static Pressure Compensation‌: Required if calibration pressure ≠ operating pressure (use manufacturer’s correction formula).
  • ‌ESD Protection‌: Always ground test equipment before HART communication.
  • ‌Data Retention‌: Store calibration logs for ≥5 years per ISO 9001.

 ‌Troubleshooting Calibration Failures‌

‌Symptom‌ ‌Potential Cause‌ ‌Resolution‌
Zero drift Diaphragm contamination Clean with IPA; replace seal
Nonlinearity Sensor damage Perform sensor characterization
HART no response Cable impedance mismatch Use 250Ω resistor in series

EL3020 PDF 

EL3020

·····································································································································································································································································································································

Our advantages:
1. The brand new product comes with a one-year comprehensive warranty.
2. All products undergo strict quality inspections.
3. Contact us and our team will provide competitive quotes and procurement plans.

 

 

Customer Reviews

0 reviews
0
0
0
0
0

There are no reviews yet.

Be the first to review “EL3020 ABB Industrial Continuous Gas Analyzer”

Your email address will not be published. Required fields are marked *