Calibration Procedure for EL3020 Transmitter
1. Pre-Calibration Setup
Tools Required:
- Reference pressure source (accuracy ≤ ±0.1% FS)
- Digital multimeter (resolution 0.01 mA)
- HART communicator / Field calibration software
Safety Precautions:
- Isolate process pressure and vent the transmitter.
- Verify ambient temperature: 15–25°C (outside range requires compensation).
2. Zero & Span Adjustment
(Apply using calibration rig)
Step | Action | Verification |
---|---|---|
1 | Apply 0% pressure (vent to atmosphere). | Output = 4.00 mA ±0.02 mA. |
2 | Adjust ZERO screw/HART parameter until output stabilizes at 4.00 mA. |
– |
3 | Apply 100% pressure (e.g., 10 bar). | Output = 20.00 mA ±0.03 mA. |
4 | Adjust SPAN screw/HART parameter to achieve 20.00 mA. |
– |
5 | Repeat Steps 1–4 until both endpoints meet tolerance. | Deviation ≤ ±0.05% FS. |
Note: For smart transmitters, use Endress+Hauser FieldCare or 375 HART Communicator to automate adjustments.
3. Linearity Verification
Test Points: 0%, 25%, 50%, 75%, 100% of range.
- Record output current at each point.
- Acceptance Criteria:Error=∣ActualmA−ExpectedmA∣SpanmA×100%≤0.1%Error=SpanmA∣ActualmA−ExpectedmA∣×100%≤0.1%
4. Post-Calibration Actions
- Seal Adjustment Screws with anti-tamper seal (if applicable).
- Update Calibration Certificate with:
- Date, technician ID
- Reference tool serial numbers
- As-found/as-left data.
- Functional Test: Simulate overpressure (110% FS) to confirm fault alarm activation (
HI-HI
limit).
Critical Notes
- Static Pressure Compensation: Required if calibration pressure ≠ operating pressure (use manufacturer’s correction formula).
- ESD Protection: Always ground test equipment before HART communication.
- Data Retention: Store calibration logs for ≥5 years per ISO 9001.
Troubleshooting Calibration Failures
Symptom | Potential Cause | Resolution |
---|---|---|
Zero drift | Diaphragm contamination | Clean with IPA; replace seal |
Nonlinearity | Sensor damage | Perform sensor characterization |
HART no response | Cable impedance mismatch | Use 250Ω resistor in series |